SPECIFICATIONS DETAILS
Specification
|
EMF(mv)
|
100℃
|
200℃
|
400℃
|
600℃
|
800℃
|
NiCr(EP)
|
2.784~ 2.844
|
5.938~ 6.002
|
12.709~ 12.819
|
19.537~ 19.699
|
26.102~ 26.308
|
|
2.814
|
5.97
|
12.764
|
19.618
|
26.205
|
CuNi(EN)
|
3.467~ 3.543
|
7.410~ 7.492
|
16.109~ 16.255
|
25.362~ 25.588
|
34.664~ 34.960
|
|
3.505
|
7.451
|
16.182
|
25.475
|
34.812
|
NiCr-CuNi
|
6.251~ 6.387
|
13.348~ 13.494
|
28.818~ 29.074
|
44.899~ 45.287
|
60.766~ 61.268
|
|
6.319
|
13.421
|
28.946
|
45.093
|
61.017
|
Physical
parameters
Specification
|
Density (g/cm3)
|
MP(℃)
|
TS(mpa)
|
Extensibility
|
Resistivity (20℃μΩ.m)
|
NiCr(EP)
|
8.5
|
1427
|
≥490
|
≥10
|
0.71
|
CuNi(EN)
|
8.5
|
1220
|
≥390
|
≥25
|
0.5
|
QUALITY GUARANTEE
SICC can supply type E thermocouple
wire material meet the accuracy standard IEC60584 and ASTM E230. The advanced melting
skills combined controls the chemical composition to achieve the highest
quality products , our produce base has the capacity of 900 tons annual with
35000 square meters that can meet your comprehensive requirements. Whatever the
bright annealed or oxidized conditions, all thermocouple conductors are
calibrated at our lab ,this includes the testing of EMF, chemical
composition, physical and thermoelectric properties before shipment.
SICC provides calibration results for
all bare thermocouple wires with each shipment, including EMF values, accuracy,
batch number, raw materials chemistry, lab environment condition,etc. Each
order has only one job number for tracement. Calibration capability is available up
to 1000C on suitable thermocouple types.
Our lab routinely calibrates at the
following fixed points:100C,200C,400C,600C,800C for type E thermocouple materials, we work continuously to
reduce the measurement uncertainty at each calibration temperature and new
estimates are made every six months to ten months, ensuring any deviations are
identified early so that results generated by the lab remain consistent and
accurate.
Contact us today for your any
requirements!